AUTOCLAVE INDICATOR STEAM 1262 100'S
Unit Of Sale: PACK/100
Manufacturers Code: 1262
Medent Code: 84500
Brand: 3M
Description
3M Attest Biological indicators that monitor the effectiveness of the steam sterilization process and provide results within 48 hours. The self-contained BI for 121℃ gravity and 132℃ vacuum-assisted steam cycles consist of a Geobacillus stearothermophilus spore strip; sealed glass ampule with growth medium and a bromocresol purple pH indicator system; brown colour-coded cap with holes for sterilant penetration and a hydrophobic filter as a bacterial barrier, and a chemical indicator on the label that changes from rose to brown when processed. After sterilization, the vial is "crushed" to join the growth media with the processed spore strip. The BI is incubated for 48 hours for a visual change readout. A color change to yellow indicates surviving spores and a positive result.
- Biological indicator vials are easy to use and interpret with visual colour readout in 48 hours.
- Self-contained biological indicators significantly reduce possibility of contamination, minimizing
false positives and assuring more accurate results.
Suggested Applications
Monitoring 121℃ gravity and 132℃ vacuum-assisted steam sterilization processes.
|